ISO 1133 ASTM D 1238 Melt Flow Indexer MFR MVR Test For PP and PE

SO 1133 ASTM D 1238 Melt Flow Indexer MFR MVR Test For PP and PE

Discription:

This is for high-precision melt flow rate instrument tester, used for testing all kinds of plastic, resin in the state of viscous flow through a certain temperature and load, every 10 min module and melt flow rate through the standard port MFR/MVR value, it is suitable for high temperature of polycarbonate, aromatic sulfone, fluorine plastics, nylon engineering plastics, can also be applied to polyethylene (PE), polistiren (PS), polipropilenă (PP), ABS rășină, polyformaldehyde (POM), policarbonat (PC) resin with lower melting temperature, such as plastic test, utilizat pe scară largă în producția de plastic, produse din plastic, petrochemical industry and related colleges and universities.

 

ISO 1133 ASTM D 1238 Melt Flow Indexer MFR MVR Test For PP and PE

Standard:

GB/T3682-2018,ISO1133-2011, ASTM D1238-13

 

Product Character:

Ø 1.More paragraphs (3) temperature control method, three groups of platinum resistance sensor, the three heating jacket separate temperature control, high accuracy intelligent PID control, ramp rate the heating speed is adjustable

Ø 2.Resolution temperature was 0.1 ℃, the temperature fluctuation of not more than 0.5 ℃ within 24 ore, the displacement sensor resolution is 0.001 mm.

Temperature and temperature protection device

Ø 3.Hand automatic cutting material, can be arbitrary set blanking time (0 ~ 999), blanking times (0 ~ 999)

Ø 4.Test can choose quality standard, or volume method, two test methods or simultaneously, and can calculate flow rate ratio.Both methods start testing when the piston rod moves to a certain location.

Ø 5.Can be set up in 240 secunde (GB3682 regulations) material preheat second countdown, after the end of the countdown, farmar automatically loaded, loaded to interval start automatic cutting, can also set the American standard (ASTM) preheating time, etc.

Ø 6.Built-in many plastic material melt index test conditions, convenient customer used to be obtained at any time, reduce the tedious work, and storing the last test conditions and results

The internal microtyping machine is convenient for printing and testing data

Ø 7.High precision touch screen controller, 7 ecran tactil inch, can be stored and print test conditions and data recently, test print menu Settings include test time (the default), the material name, condiții de test (temperatura, weight weight, blanking intervals), test results data.

Ø 8.Reserved RS232 interface or USB interface to connect computers, can through the computer control instrument operation and to save test data transmission to the computer,

Ø 9.Can manually loading weight or by stepping motor drives the ball screw rotation in order to achieve the purpose of automatic control at the top of the load weight increase or decrease

 

Test Methods:MFR +MVR Methods

Modul de afișare:English color touch screen display

 

ISO 1133 ASTM D 1238 Melt Flow Indexer MFR MVR Test For PP and PE

Specificație

 

 

Model LR-A001-C
Barrel parameters Inner hole 9.550±0.005mm length 163mm
Pistons head parameters Piston head: 9.474±0.005mm
Piston rod diameter 8.94mm
Mold length 8.000±0,025 mm
Mold nominal inner diameter 2.0955±0.0051mm

 

Half die length 4.000±0,025 mm
Half die nominal inner diameter 1.050±0.005mm
Temperature parameter With intelligent temperature control instrument, with four pairs of significant temperature setting control, PID parameters can be automatically set, precision up to ± 0.1 degree centigrade
Interval de temperatură Room Temperature~500degree centigrade
fluctuație de temperatură ±0.2 ℃
Uniformitatea temperaturii ± 1 ℃
Rezoluția afișajului temperaturii 0.1℃
rezoluţie afișaj 0.1℃
Maximum comsuption < 600W
Rezoluția afișajului timpului 0.1s
Weight parameters are as follows:
Precizia greutății ± 0.5%
Configurare de bază

 

A 0.325kg
B 1.2kg
C 2.16kg
D 3.8kg
E 5kg
 
F 10kg
 
G 12.5kg
H 21.6kg
 
Position detection
Loop distance from up and down 30mm
Control precision ± 0.1mm
Test flow control
Times of cutting the material 0~10 times
Material cutting interval 0~999s(set reference Table 2)
Control flow reaches the set temperature without volatility
Barrel temperature time 15 mă.
Material be installed 1 mă.
Material sample temperature recovery time 4 mă.
When binder set 1mă
Output mode

Micro-automatic print output

Cut material methods The hand automatic cut into one body
Testul de sarcină Eight sets of weights